TEST LABS
Moisture Tester for Insulating Liquids — Coulometric Karl Fischer
The HVEX Moisture Tester determines the water content of insulating oils and other liquids by the coulometric Karl Fischer method, suited to very small amounts of water — the typical range of in-service transformer oil. The sample is injected into the titration cell; the electrolysis current generates iodine at the anode, which reacts with water in the presence of sulfur dioxide, base and methanol. Platinum electrodes detect the end point, and the water mass is calculated from the applied charge, by Faraday’s law. With a 3 µg to 100 mg range and 0.1 µg resolution, the instrument corrects the reagent blank and equilibrium drift, applies 9 calculation formulas and converts a single injection to ppm, mg/kg, mg/L or %. History is saved with date and time, and RS232 and RS485 ports send results to a computer. Compact (4.5 kg) and low-power, it serves utility, maintenance, chemical and pharmaceutical laboratories.