HVEX

Command Palette

Search for a command to run...

HVEX | Moisture Tester for Insulating Liquids — Coulometric Karl Fischer

TEST LABS

Moisture Tester for Insulating Liquids — Coulometric Karl Fischer

The HVEX Moisture Tester determines the water content of insulating oils and other liquids by the coulometric Karl Fischer method, suited to very small amounts of water — the typical range of in-service transformer oil. The sample is injected into the titration cell; the electrolysis current generates iodine at the anode, which reacts with water in the presence of sulfur dioxide, base and methanol. Platinum electrodes detect the end point, and the water mass is calculated from the applied charge, by Faraday’s law. With a 3 µg to 100 mg range and 0.1 µg resolution, the instrument corrects the reagent blank and equilibrium drift, applies 9 calculation formulas and converts a single injection to ppm, mg/kg, mg/L or %. History is saved with date and time, and RS232 and RS485 ports send results to a computer. Compact (4.5 kg) and low-power, it serves utility, maintenance, chemical and pharmaceutical laboratories.

Technical Information

Power supply
220 VAC ± 20%
Reagent
Coulometric Karl Fischer reagent suitable for the cell
Sampling
Clean, dry syringe; balance to weigh the sample
Environment
5 to 40 °C and relative humidity below 80%
Bench
Stable surface, free of vibration and excess humidity

Water measured in micrograms

Iodine is generated electrolytically in the cell itself: each milligram of water corresponds to 10.71 C of charge. That makes the method absolute, with no titer standardization, and accurate for the low water contents of insulating oil.

HVEX | Water measured in micrograms
HVEX | Benefits and Usability

Benefits and Usability

  • HVEX | High sensitivity
    High sensitivity

    Range from 3 µg to 100 mg of water, with 0.1 µg resolution.

  • HVEX | Results in several units
    Results in several units

    One injection converted to ppm, mg/kg, mg/L or %, saving reagent.

  • HVEX | Automatic corrections
    Automatic corrections

    Reagent blank and equilibrium drift compensation.

  • HVEX | Records and communication
    Records and communication

    History with date and time and RS232 and RS485 ports.

Tester highlights

Water content in insulating oil with laboratory accuracy.

  • HVEX | ImageCoulometric Karl Fischer
  • HVEX | Image0.1 µg resolution
  • HVEX | ImageColor LCD
  • HVEX | ImageFor transformer oil